Ajou University repository

Finite element analyses of lateral condyle fracture fixation in paediatrics regarding configuration of Kirschner-wireoa mark
  • Jeon, Sangbin ;
  • Ahn, Wooyeol ;
  • Oh, Jongbeom ;
  • Chung, Jaiwoo ;
  • Choi, Junwon ;
  • Lee, Soonchul
Citations

SCOPUS

1

Citation Export

Publication Year
2022-12-01
Publisher
BioMed Central Ltd
Citation
BMC Musculoskeletal Disorders, Vol.23
Keyword
Finite element analysesFixationKirschner-wireLateral condyle fracture
Mesh Keyword
Bone WiresChildFinite Element AnalysisFracture FixationFracture Fixation, InternalHumansHumeral FracturesHumerusMalePediatrics
All Science Classification Codes (ASJC)
RheumatologyOrthopedics and Sports Medicine
Abstract
Background: This study aimed to discover the most stable outcome among different Kirschner-wire (K-wire) configurations for fixation of a lateral condyle fracture (Milch type II) in different loads of stress by using finite element analyses (FEA). Methods: The right humerus of a 6-year-old boy with a lateral condyle fracture (Milch type II), was modelled with a computer aided engineering. Using FEA, peak von Mises stress and stiffness were evaluated first for a single K-wire fixation by varying the angle (0, 5, 10, 15, 20, 25, 30 degrees). Then, based on the single K-wire result, assessment of peak von Mises stress and stiffness were evaluated via FEA for two- or three-wire fixation under various configurations (two convergent, two parallel, three divergent). Results: Single K-wire fixation by 5 and 25 degrees had the lowest peak von Mises stress. The fracture site showed higher stiffness at 0, 5 and 15 degrees. Considering the collected results and clinical situation, 5 degree K-wire was selected for the FEA of multiple K-wire fixation. For multiple K-wire fixation, three divergent (5–20-35 degrees) K-wires showed better stability, both in peak von Mises stress and stiffness, than any two-K-wire configurations. Among two K-wire fixations, two divergent (5–50 degrees) K-wires provided the lowest von Mises stress in varus and valgus while two divergent (5–65 degrees) K-wires showed better results in flexion, extension, internal and external rotation, and both configurations showed similar results in stiffness. Conclusions: We successfully created a paediatric lateral condyle fracture (Milch type II) model which was used to conduct FEA on different K-wire configurations to achieve stability of the fracture. Our results show that an initial K-wire inserted at 5 degrees, followed by the insertion of a second divergent wire at either 45 or 60 degrees provides the most stability in two K-wire fixations in this type of fracture repair.
ISSN
1471-2474
Language
eng
URI
https://dspace.ajou.ac.kr/dev/handle/2018.oak/33021
DOI
https://doi.org/10.1186/s12891-022-05897-3
Fulltext

Type
Article
Funding
This work was supported by the National Research Foundation of Korea (NRF) and a grant funded by the Korean government (MSIT) (No. 2022R1A2C2005916).
Show full item record

Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.

Related Researcher

Choi, Jun Won Image
Choi, Jun Won최준원
College of Bio-convergence Engineering
Read More

Total Views & Downloads

File Download

  • There are no files associated with this item.