Citation Export
DC Field | Value | Language |
---|---|---|
dc.contributor.author | Kim, Ka Ram | - |
dc.contributor.author | Lee, Kyung Won | - |
dc.contributor.author | Chun, Hyeong Jin | - |
dc.contributor.author | Lee, Danbi | - |
dc.contributor.author | Kim, Jae Ho | - |
dc.contributor.author | Yoon, Hyun C. | - |
dc.date.issued | 2022-01-15 | - |
dc.identifier.uri | https://dspace.ajou.ac.kr/dev/handle/2018.oak/32336 | - |
dc.description.abstract | Herein, we introduce a smartphone-integrated immunosensor based on non-spectroscopic optical detection. Sedimentation of the retroreflector and gentle inversion of the microfluidic chip was chosen as biosensing principles to ensure minimal human involvement. To realize this, wash-free immunosensing was implemented on a polymeric microfluidic chip device fabricated for light signal penetration in retroreflection signal acquisition. Applying a transparent chip and passive modulation of retroreflectors enabled the minimization of human error during sensing. In addition, a retroreflection-detectable optical gadget was constructed for integration with the commercial smartphone. The gadget had an optical chamber that induced retroreflection by integration with a smartphone. When the micro-sized reflector, named the retroreflective Janus microparticle, reacted on the sensing surface, the incident light was retroreflected towards the image sensor and quantified by a smartphone-installed Android application package. The developed application package features include time-lapse image capture performed by manipulating LED flash and camera modules, and quantification of retroreflected signal counts by image processing of time-lapse images. With this platform, the user could independently commence optical signal processing without a complicated optical setup and running software on a PC, and sensitive and reproducible immunosensing results could be obtained. The applicability test for creatine kinase-myocardial band detection from the buffer to serum was conducted and presented a calibration curve of 0–1000 ng/mL within 1 h. With the developed system, we believe that the applicability of the platform in bioanalytical detection can be expanded. | - |
dc.description.sponsorship | This work was supported by the Creative Materials Discovery Program (NRF- 2019M3D1A1078943 ), the Priority Research Centers Program (NRF- 2019R1A6A1A11051471 ), and the SGER Program (NRF- 2018R1D1A1A02085354 ) funded by the National Research Foundation of Korea . HCY also acknowledges the support from the Samsung Research Funding Centre of Samsung Electronics . | - |
dc.description.sponsorship | This work was supported by the Creative Materials Discovery Program (NRF-2019M3D1A1078943), the Priority Research Centers Program (NRF-2019R1A6A1A11051471), and the SGER Program (NRF-2018R1D1A1A02085354) funded by the National Research Foundation of Korea. HCY also acknowledges the support from the Samsung Research Funding Centre of Samsung Electronics. | - |
dc.language.iso | eng | - |
dc.publisher | Elsevier Ltd | - |
dc.subject.mesh | Creatine kinase | - |
dc.subject.mesh | Creatine kinase-MB | - |
dc.subject.mesh | Immunosensing | - |
dc.subject.mesh | Micro particles | - |
dc.subject.mesh | Microfluidic-chips | - |
dc.subject.mesh | Optical bio-sensing | - |
dc.subject.mesh | Retro reflection | - |
dc.subject.mesh | Retro-reflectors | - |
dc.subject.mesh | Smart phones | - |
dc.subject.mesh | Smartphone-based biosensor | - |
dc.subject.mesh | Biosensing Techniques | - |
dc.subject.mesh | Humans | - |
dc.subject.mesh | Immunoassay | - |
dc.subject.mesh | Lab-On-A-Chip Devices | - |
dc.subject.mesh | Microfluidics | - |
dc.subject.mesh | Smartphone | - |
dc.title | Wash-free operation of smartphone-integrated optical immunosensor using retroreflective microparticles | - |
dc.type | Article | - |
dc.citation.title | Biosensors and Bioelectronics | - |
dc.citation.volume | 196 | - |
dc.identifier.bibliographicCitation | Biosensors and Bioelectronics, Vol.196 | - |
dc.identifier.doi | 10.1016/j.bios.2021.113722 | - |
dc.identifier.pmid | 34700265 | - |
dc.identifier.scopusid | 2-s2.0-85117695145 | - |
dc.identifier.url | www.elsevier.com/locate/bios | - |
dc.subject.keyword | Creatine kinase-MB | - |
dc.subject.keyword | Immunosensing | - |
dc.subject.keyword | Optical biosensing | - |
dc.subject.keyword | Retroreflection | - |
dc.subject.keyword | Smartphone-based biosensor | - |
dc.description.isoa | false | - |
dc.subject.subarea | Biotechnology | - |
dc.subject.subarea | Biophysics | - |
dc.subject.subarea | Biomedical Engineering | - |
dc.subject.subarea | Electrochemistry | - |
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.