Ajou University repository

비스무스를 이용한 방사선 피폭 감소 및 화질 연구
  • 김재석
Citations

SCOPUS

0

Citation Export

Advisor
김영길
Affiliation
아주대학교 일반대학원
Department
일반대학원 의용공학과정
Publication Year
2020-02
Publisher
The Graduate School, Ajou University
Description
학위논문(박사)--아주대학교 일반대학원 :의용공학과정,2020. 2
Alternative Abstract
By applying an ergonomically developed Bismuth shield to the, endovascular treatment of cerebral aneurysm the radiation dose of the scalp and lens from the medical radiation exposure was reduced. The enrtance surface dose was analyzed by measuring the occipital parts, bilateral temporal parts, bilateral quadriceps, and nasal tip of the developed bismuth shield using a photostimulable fluorescence dosimeter before (Group A) before use (Group B). Signal to noise ratio (SNR) and contrast to noise ratio (CNR) analysis were used to evaluate the image quality when Bismuth shielding was used. The mean entrance surface dose of A group and B group was 26.92% lower than that of A group. The analysis of CNR and SNR was the same for both Roadmap and DSA. The use of Bismuth shielding is an alternative that can reduce the radiation impairment due to temporary hair loss and other stochastic effects that may occur after cerebrovascular intervention.
Language
eng
URI
https://dspace.ajou.ac.kr/handle/2018.oak/20802
Fulltext

Type
Thesis
Show full item record